Building an uncertainty budget for a standard solution: from certificate to mg/mL

Reference standards & documentation Published By RSSE Technical Team · Reviewed by RSSE Quality Unit

The uncertainty on a certificate covers the material in the vial, not the solution made from it. This works through a full budget for a 25 mg in 25 mL preparation: three components, one result, and the two places where a correct-looking calculation still understates the error.

On this page

The certificate covers the vial, not the solution

A certificate states an assigned value and an expanded uncertainty for the material as supplied: that solid, in that container, under the storage it names, at the time it was characterised. Weigh a portion and dissolve it and you have created a different quantity, the concentration of your solution, which carries its own uncertainty. It is never smaller than the certificate's, and it is usually larger by an amount that depends on your balance and your glassware rather than on the producer.

The failure this prevents is specific. A laboratory copies the plus or minus 0.50 per cent onto the label of a 1 mg/mL working solution. On the preparation below the correct value is 0.58 per cent at the same coverage factor; from a 5 mg portion it is 1.9 per cent. The certificate cannot say which applies, because it does not know how the portion was taken.

The model, in relative terms

For a standard weighed and made to volume, C = m x P / V, with m the mass in milligrams, P the assigned mass fraction as a decimal, V the volume in millilitres and C the result in mg/mL. Multiplication and division mean the relative uncertainties add in quadrature:

(u(C)/C)^2 = (u(m)/m)^2 + (u(P)/P)^2 + (u(V)/V)^2

This is the root-sum-of-squares combination set out in the GUM, and it holds while the three inputs are independent. That condition is where most budgets quietly break.

Component one: the mass taken

Four inputs, only one of which is on the certificate.

InputAt a 25.00 mg portionEvaluationStandard uncertainty
Balance calibration, U = 0.04 mg at k = 20.020 mgType B0.020 mg
Repeatability, s = 0.015 mg from ten weighings-Type A, 9 df0.015 mg
Buoyancy, left uncorrected at 0.09 per cent0.0225 mgType B, rectangular0.0130 mg
Display resolution, 0.01 mg-Type B, rectangular0.0058 mg
Combined--0.0288 mg, 0.115 per cent

Buoyancy is the term usually left out. A balance reports conventional mass, defined against weights of density 8.0 g/cm3 in air of 1.2 mg/cm3. An organic solid near 1.3 g/cm3 occupies more volume than the weights balancing it and has a true mass about 0.09 per cent above the reading. Left uncorrected it is a bias; entered as a rectangular distribution it is a small term; treating the reading as exact is neither. The definition is in NISTIR 6969, and the weighing step carries its own minimum-weight arithmetic.

Component two: the assigned value

The largest term, and the one most often misread. Two distinctions decide it.

Percentage points, or per cent of the value. A certificate reading 99.30 per cent with an expanded uncertainty of 0.50 per cent may mean 0.50 points on the value or 0.50 per cent of it. At 99.30 per cent the difference is 0.0035 points and does not matter; at 50 per cent it is a factor of two.

As supplied, or dried. A value assigned on the material as received includes the water it holds. A value assigned after drying at a stated condition applies to dried material, so weighing as received requires you to apply the loss on drying yourself, which adds a component you must measure. The two bases are not interchangeable, and on a hygroscopic salt the gap reaches 0.5 per cent, larger than every other term here.

Here P = 0.9930 with U = 0.50 points at k = 2, so u(P) = 0.0025 and u(P)/P = 0.252 per cent.

One thing in this slot is not a component: a chromatographic area percentage. Peak-area normalisation is not a mass fraction, so putting it where P belongs introduces a bias of unknown size. Mass balance or qNMR is what produces a value that can occupy this position.

Component three: the volume

A one-mark flask contributes three things: the tolerance it was made to, the temperature you filled it at against the temperature its capacity is defined at, and how reproducibly you stop at the mark.

InputAt 25 mL, Class AEvaluationStandard uncertainty
Tolerance marked on the flask, 0.03 mL0.0300 mLType B, rectangular0.0173 mL, 0.069 per cent
Temperature, 20 plus or minus 4 degrees C, 0.00021 per degree0.0210 mLType B, rectangular0.0121 mL, 0.048 per cent
Repeatability of filling to the mark, s = 0.010 mL-Type A, 9 df0.0100 mL, 0.040 per cent
Combined--0.0234 mL, 0.093 per cent

The tolerance deserves a caution. ISO 1042 defines the capacity of a one-mark flask at 20 degrees C and sets two accuracy classes, so the figure stamped on the glass is an upper bound on the error of any flask that passed, not the error of your flask. ISO 4787 exists because that difference matters: it gives methods for testing an individual instrument and determining its error at one or more points. A tested flask carries a smaller and defensible Type B input; an untested one carries the class tolerance.

Temperature is often dismissed and should not be. Water expands by roughly 0.00021 per degree C, so a flask filled at 28 degrees C holds about 0.17 per cent more volume than one filled at 20 degrees C, some two and a half times the tolerance term.

Combining, and what the total says

ComponentRelative standard uncertaintyVarianceShare of variance
Mass taken0.115 per cent0.013316 per cent
Assigned value0.252 per cent0.063474 per cent
Volume0.093 per cent0.008710 per cent
Combined0.292 per cent0.0854100 per cent

Combined relative uncertainty 0.292 per cent, so U = 0.58 per cent at k = 2 and the solution is 0.9930 plus or minus 0.0058 mg/mL.

The shares matter more than the total. The assigned value carries 74 per cent of the variance, the weighing 16 per cent and the flask 10 per cent. A five-place balance and a tested flask have taken the two smaller terms close to their floor. What remains cannot be improved by any balance because it is the certificate: when a method needs a solution better than 0.6 per cent, the lever is a tighter certificate, not a better laboratory.

One pipette, two deliveries

Take a 1:100 dilution done as two 1:10 steps with the same 10 mL pipette. It is tempting to enter the pipette's 0.07 per cent twice and combine in quadrature for 0.10 per cent. That is wrong. The calibration error of one physical pipette is the same error in both steps: if it delivers 0.07 per cent high the first time it does so the second time, and the dilution factor is biased by 0.14 per cent rather than by a random quantity near 0.10. Correlated components add linearly, and the quadrature sum applies only to independent ones.

The test is whether both deliveries come from the same physical instrument. Two pipettes of the same class have independent errors and combine in quadrature; one pipette used twice does not.

Bias is not uncertainty

The GUM requires a correction for every recognised systematic effect, and the uncertainty entered in the budget is the uncertainty of that correction rather than the size of the effect. Four things here are corrections, and entering them as random terms hides them:

  • Using a peak-area percentage as the assigned value. Correct it by using an assay value, or do not claim a mass fraction.
  • Ignoring the salt or hydrate factor when a concentration is stated on the free base. The conversion factors belong in the equation, not in the uncertainty.
  • Drying a standard whose certificate says use as is. That alters the material, not the uncertainty.
  • Filling at 28 degrees C and calling the volume 25.00 mL. Control the temperature, or carry the volume at the temperature actually used.

When k = 2 is not enough

Expanded uncertainty is the combined standard uncertainty multiplied by a coverage factor, and k = 2 corresponds to about 95 per cent only when the combined uncertainty is approximately normal and its own uncertainty is negligible. That second condition is a statement about degrees of freedom. Where a large Type B term dominates, the effective degrees of freedom are effectively infinite and k = 2 is sound. Where a Type A term from a handful of replicates dominates it is not: at eight effective degrees of freedom the factor for 95 per cent is 2.3 rather than 2.0, and a budget reporting k = 2 claims coverage it does not have. In a standard-solution budget the Type A terms are usually the small ones, so k = 2 is normally defensible.

What a gravimetric preparation changes

Weighing the diluent instead of filling to a mark removes the volumetric component: 25 g of diluent on a five-place balance carries a relative uncertainty near 0.001 per cent, against 0.093 per cent for the flask. It also changes the unit to mg/g, so the density of the solution never enters the budget, which is why gravimetric reference materials are commonly certified in mg/kg.

What it does not remove is the dominant term. The total falls only from 0.58 to 0.55 per cent, because the same 74 per cent of the variance is still the certificate. A gravimetric preparation buys robustness, not a smaller number.

The budget has a date

Every figure above describes the solution on the day it was made. A stored solution acquires a stability component from adsorption to the container wall, evaporation through a septum or hydrolysis of the analyte, and that component is on no certificate, because the producer characterised a solid and not your liquid. Where a hold time has been established it enters as a Type B term from the study; where it has not, the honest statement is that the uncertainty after storage is unknown. Establishing a hold time is the measurement that converts the second case into the first.

Five questions for a supplier

  • Is the stated uncertainty expanded, and at what coverage factor?
  • Is the assigned value on an as-supplied or a dried basis, and what is the drying condition?
  • Is the purity a mass fraction from an assay, or a chromatographic area percentage?
  • Does the certificate state metrological traceability, and to what?
  • What is the smallest unit, and does the assigned value apply to the whole unit or to a portion?

ISO 17034:2016 is the standard a producer is assessed against for competence and consistent operation, and it covers the production of all reference materials including certified ones. A certificate that cannot answer them is not evidence that the material is wrong; it is evidence that the uncertainty of your solution cannot be calculated from it.

Frequently asked

Can I use the certificate uncertainty as the uncertainty of my solution?
No. The certificate describes the material as supplied. A solution adds the weighing, the volumetric step and the preparation itself, so its uncertainty is at least the certificate value and usually larger. In the worked example the 0.50 per cent on the certificate becomes 0.58 per cent for a 1 mg/mL solution at the same coverage factor.
Which term dominates a standard solution made from a certified material?
The assigned value, in most cases. In the worked example it carries 74 per cent of the variance, against 16 per cent for the weighing and 10 per cent for the flask. That is why a better balance does not improve the solution much once the two smaller terms are near their floor.
Is 0.5 per cent uncertainty on a certificate absolute or relative?
It depends on the certificate wording, and the two are not the same. On a 99.30 per cent assigned value, 0.50 percentage points and 0.50 per cent of the value differ by 0.0035 points. On a 50 per cent assigned value they differ by a factor of two, and the budget changes with it.
Do two dilutions with the same pipette combine in quadrature?
Not for the pipette calibration component. One physical pipette carries the same calibration error into both steps, so that component adds linearly: two 0.07 per cent deliveries bias the dilution factor by 0.14 per cent, not by the 0.10 per cent a quadrature sum gives. Independent instruments of the same class do combine in quadrature.
Is k = 2 always the right coverage factor?
Only when the combined uncertainty is approximately normal and its own uncertainty is small, which in practice means a large effective number of degrees of freedom. At eight effective degrees of freedom the factor for about 95 per cent is 2.3, so a budget dominated by a Type A term from a few replicates should not report k = 2.
Does a gravimetric preparation eliminate the volumetric uncertainty?
It removes it and replaces the unit. Weighing the diluent takes the volumetric term from 0.093 per cent to about 0.001 per cent, and the result is then a mass fraction in mg/g rather than a concentration in mg/mL, so solution density never enters the budget. The certificate term remains dominant and the total falls only from 0.58 to 0.55 per cent.

Sources and further reading

Every document below is published by the body that owns it; none of them is a summary written by us. Accessed 2026-11-13.

Send us a CAS number

Availability, documentation and lead time — answered by the people who handle the batch, not a ticket queue.

Request a quote sales@rssechem.com

How to read a certificate of analysis Field by field: identity, purity, content basis, the retest date, and the questions worth asking a supplier before you file the document.How to choose a reference standard supplier A practical checklist for evaluating reference standard suppliers on documentation, traceability, lot consistency, logistics and technical support.Stable isotope-labelled standards and when to use them When labelled internal standards earn their place in an LC-MS or GC-MS method, what to specify when ordering, and the traps in quantitative work.Nitrosamine and NDSRI standards: what to test for and what to buy The difference between small nitrosamines and NDSRIs, what a laboratory needs as a standard for each, and how to specify one so the result stands up to review.What is an impurity reference standard? What the term means, how it differs from a pharmacopoeial standard and a working standard, what the documentation has to state, and how to pick the right form and purity basis.ICH M7 in practice: classifying mutagenic impurities How the five-class hazard assessment works, what the 1.5 microgram default really covers, and how the four control options decide which standards you buy.qNMR vs mass balance: how purity is really assigned Two certificates, one compound, two purity values. What mass balance and qNMR each measure, why the figures differ, and how the basis changes the number.Elemental impurities: from a PDE to a limit you can measure How a permitted daily exposure becomes a concentration limit, why the dose moves the number, and what an elemental impurity specification must state.Salt, hydrate and free base: converting content so the numbers reconcile How to move content and impurity values between as-is, anhydrous and free-base bases, in what order, and why the basis can turn a pass into a fail.Forced degradation study design: how far to push each stressor How to set stressor strength, read a kinetics screen, judge mass balance and catch over-degradation before it wastes a method development cycle.Analytical method transfer: acceptance criteria you can defend Four transfer routes, criteria built from the method's own precision, and the standard's assigned content that shifts every result by a fixed amount.Chiral and stereoisomeric impurity standards Enantiomer, diastereomer, epimer, E/Z isomer: four different standards and four different questions. What each has to state before it can control a method.Reference standard, analytical standard, CRM: what the words mean Four labels appear on the same vial in different catalogues. They are not synonyms, and the difference decides what you may claim in a report.Peptide reference standards: what changes when the analyte is a chain Deletion sequences, epimers, pyroglutamate, deamidation, counter-ion and water: why a peptide standard needs a different certificate from a small molecule.Cleaning validation residue limits: where the number comes from How a residue limit is derived from the dose, from 10 ppm and from a health-based PDE, and what the swab method and its standard must be able to reach.Cold chain and stability of reference standards: what transport and receipt checks protect Storage regimes, stability budgets and mean kinetic temperature, plus the receipt checks that decide whether a standard still supports its certificate.Impurity profile comparison: what a reference standard proves, and what it does not How reporting, identification and qualification thresholds set the scope of an impurity profile comparison, and where a standard stops proving anything.Weighing a reference standard: minimum weight and moisture How the 0.10% repeatability rule sets a minimum weight, why a hygroscopic standard is used as is, and which figures belong on the worksheet.Standard solution stability: how long a prepared standard can be used How to set a hold time for a standard solution: two baselines, six failure modes, a defensible study design and criteria fixed before the data arrive.Reporting, identification and qualification thresholds by dose How the 0.05, 0.10 and 0.15 per cent thresholds are banded by dose, where the milligram cap takes over, and what each limit forces you to test.USP 233 in practice: the J value and the four preparation routes How chapter 233 turns a limit into a J value, which of four preparation routes to use, and the acceptance criteria a method must meet before routine use.